热管论坛  联系我们  收藏本站  汇款信息  咨询回复  热管论文  
您现在的位置: 主页 >> 热管技术 >> 辅助知识
等离子光谱仪最佳分析参数的研究

日期:2008-4-10 0:12:12 来源:来自网络 查看:[ ] 作者:成 勇 肖 军 宁燕平 热度:

摘 要 采用新型等离子体光谱仪(IRIS/HR型),对同时测定碳钢、低合金钢中Si,Mn,P,V,Ti,Cu,Co,Ni,Cr,Mo,Mg,Al12个元素的最佳分析参数进行了试验。研究了分析谱线、背景校正点、高频功率、载气压力、试样提升量和辅助气体流量对分析结果的影响。将试验所选择的分析参数应用于标准样品的分析,结果表明该方法简便易行、快速、准确。
关键词 等离子光谱仪 碳钢 低合金钢 分析参数

INVESTIGATION OF OPTIMUM ANALYTICAL PARAMETERS
FOR PLASMA OPTICAL SPECTROMETER

Cheng Yong Xiao Jun Ning Yanping
(Panzhihua Iron & Steel Research Institute)

Abstract The experiments for selecting optimum analytical parameters of simultaneous determination of twelve elements,i.e.Si,Mn,P,V,Ti,Cu,Co,Ni,Cr,Mo,Mg,Al in carbon steel and low-alloy steel were carried out using new type(IRISHR)plasma optical spectrometer.The effects of analytical spectral line,back ground correcting point,high frequence power,carrier gas pressure,flow rate of liquid sample and auxiliary gas on analytical results were investigated.The analytical parameters selected are used for analyzing standard sample.The results obtained show that the features of this method are simple,rapid and accurate.
Key Word
 plasma spectrometer,carbon steel,low-alloy steel,analytical parameter


1 引言

  近年来原子发射光谱分析技术取得了重大进展,出现了以电荷注射阵列检测器(CID)与中阶梯光栅交叉二维分光系统相结合而组成的全新概念的全谱直读等离子体发射光谱仪,它区别于早期传统意义上的以光电倍增管(PMT)作检测器的单道扫描式和多道固定式光谱仪。由于CID阵列检测器是在一块28mm×28mm的硅基上制出512×512个光敏检测单元,接收谱线时各检测单元独立工作,彼此不存在任何耦合和影响,使得在一次进样分析过程中可测定和记录样品中所有元素的所有波长的谱线。所以,在175~800nm的波长范围内可测定其中任一条谱线,方便地实现一个元素多条谱线同时测定或高含量元素用弱谱线、微量元素用灵敏谱线的分析方式。具有全谱直读和高灵敏度的特点,并能以其灵活而独特的同步扣背景功能消除光谱背景干扰[1]
  本文以同时测定碳钢、低合金钢中Si,Mn,P,V,Ti等十二个元素的分析方法为例,对这类新型全谱直读等离子体光谱仪的分析性能进行探讨,重点研究该类仪器测定钢铁样品的最佳分析测试参数,将试验所获得的参数应用于标准样品的实际分析,得到令人满意的结果。


2 仪器与试剂

2.1 仪器
  美国TJA公司IRIS/HR型全谱直读等离子体发射光谱仪,中阶梯光栅,二维电荷注射阵列检测器,175~800nm的波长范围,P200计算机并配备Thermo SPEC/CID分析软件。
2.2 试剂
  HNO3:优级纯;HCl:优级纯;
  纯铁粉:GBW01402-b;
  光谱纯的Al,Ti,Fe,MgO,K2Cr2O7,CuO,Co2O3,NiO,MoO3,V2O5,KH2PO4,SiO2,并将各元素配制成1g/L的标准溶液;
  碳钢和低合金钢标准样品。

3 仪器最佳分析参数的选择

3.1 分析谱线与背景校正点的选择
  由于不同种类的碳钢、低合金钢中各待测元素含量变化范围大,为适应不同样品的分析测定,满足待测组分有时高有时低的情况,每个元素分别选择了灵敏谱线和弱谱线3~5条,并扫描制作了每条谱线的局部阵列图谱。结合每条谱线的检出强度、信噪比要大、干扰要小等因素,进行综合分析,最后优选出了各元素的最佳分析波长及背景校正位置,见表1。

表1 各元素的分析谱线和背景校正点

元素 分析波长/

nm

阵列窗宽/
pixel
级 次 扣背景点/pixel 检 测 宽 度
pixel
左 边 右 边
Si 251.612 27 133   22 3
288.158 27 116 2 22 3
Mn 293.306 15 114   13 2
293.930 27 114 5   3
P 178.287 27 188 1 21 2
213.618 27 157 1 26 2
V 309.311 27 133 5 26 1
310.230 27 108 6 22 1
Ti 334.941 27 100 1 26 1
336.121 27 100 5 22 2
Al 309.271 27 108   21 1
396.152 27 85 4   3
Cu 324.754 27 103 7 26 1
327.396 15 102 1 14 3
Co 228.616 27 147 1 24 3
Ni 227.021 15 148   12 2
341.476 27 98 5   1
Cr 283.563 27 118 5 23 1
267.716 15 125 2 14 2
Mo 202.030 27 166 9 24 1
281.615 27 119 1 22 2
Mg 279.553 15 120 1 13 3
285.213 15 117 2   1

  根据图谱资料[2]获知表1中有的分析谱线或受基体铁峰干扰、或受共存元素影响、或为双峰现象。例如在谱线Mg(285.213nm)、Si(288.158nm)、Mo(202.030nm)的旁边就存在着基体铁峰;而213.598nm的Cu谱线对P(213.618nm)以及309.311nm的V谱线对Al(309.271nm)的测定具有旁峰干扰;测定谱线Ti(334.941nm)的旁边有Ti的另一条谱线334.904nm的干扰。为此我们用1000ppm的干扰元素对10ppm的测定元素进行了干扰试验,由于该类仪器的分辨率足以使相邻两峰分离,干扰元素在测定元素波峰处的信号强度同基线持平,不产生实际干扰,所以测定时只需保证将测定峰位准确定位于被测谱线的峰位上并选择适宜背景校正点即可。元素谱线干扰情况见图1。

60t1.gif (10058 bytes)
60t2.gif (11127 bytes)
60t3.gif (11386 bytes)
图1 元素谱线干扰图谱

3.2 选择高频发生器功率
  改变发生器功率,并保持雾化器压力、试液提升量等参数不变,使之分别在不同功率条件下进行标准样品溶液的测定。以Si,Mn,P,V,Ti为代表计算其信噪比,制作功率—信噪比曲线图,见图2。

61t1.gif (4852 bytes)
图2 高频发生器功率对元素测定的影响

  从图2可见,随着功率的增大,信噪比逐渐下降并趋于平缓。同时综合考虑到火焰的稳定性,故将高频功率设置成1150W为宜。
3.3 选择雾化器压力
  只改变雾化器压力,并保持其它测定参数不变,使之分别在不同雾化器压力条件下进行标准样品溶液的测定。以Si,Mn,P,V,Ti为代表计算其信噪比,制作雾化器压力—信噪比曲线图,见图3。

61t2.gif (5428 bytes)

图3  雾化器压力对元素测定的影响

  从图3可见:随着雾化器压力的增大,信噪比逐渐升高,但升高到一定程度后有的元素其信噪比趋于平缓,有的反而下降。综合考虑各个元素后,将雾化器压力设置为32PSI为宜。
3.4 选择试液提升量
  只改变试液提升量,并保持其它测定参数不变,分别在不同试液提升量的条件下进行标准样品溶液的测定。以Si,Mn,P,V,Ti为代表计算其信噪比,制作试液提升量—信噪比曲线图,见图4。

61t3.gif (6389 bytes)
图4 试液提升量对元素测定的影响

  从图4可见,试液提升量的变化对各元素的信噪化并无多大影响。将试液提升量设置为2.22mL/min。
3.5 选择辅助气流量
  只改变辅助气流量,并保持其它测定参数不变,分别在不同辅助气流量的条件下进行标准样品溶液的测定。以Si,Mn,P,V,Ti为代表计算其信噪比,制作辅助气流量—信噪比曲线图,见图5。

61t4.gif (4263 bytes)

图5  辅助气流量对元素测定的影响

  从图5可见,辅助气流量对元素信噪比影响不大。考虑到实际工作中样品多数为铁基,辅助气增大可减少对炬管的侵蚀,故辅助气流量设置成1.0L/min为宜。
3.6 仪器的最佳分析测试参数
  经上述优化试验,得出了仪器的最佳测试参数,见表2。

表2 仪器最佳分析测试参数

功率/

W

雾化压力/
PSI
试液提升量/
mL.min-1
辅助气流量/
mL.min-1
积分时间/s
>300nm波长段 <300nm波长段
1150 32 2.22 1.0 5 10

4 试验结果

4.1 制作标准曲线
  为了使校正标准溶液与实际试样基体相匹配,溶解0.2500g高纯铁粉(与试样称重量相当)五份于100mL容量瓶中作基体打底,然后分别加入各被测元素的标准溶液,稀释定容。以此5个点制作工作曲线。各元素含量见表3。

表3 工作曲线中各元素的含量%

  Si Mn P V Ti Cu Co Ni Cr Mo Al Mg
空白点 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.01
校正点1 0.1 0.5 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05
校正点2 0.5 1.0 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1
校正点3 1.0 1.5 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3
高点 1.6 2.0 0.6 0.6 0.6 0.6 0.6 0.6 0.6 0.6 0.6 0.6

4.2 试样分析
  称0.2500g试样于250mL锥形瓶中,加入60mL(1+5)HNO3加热溶解试样至完全,再加入10mL(1+1)HCl煮沸,冷却后移入100mL容量瓶定容、摇匀。按上述选定的分析参数进行测试。
4.3 方法的精密度
  对GBW1307、U71Mn标准样品进行8次平行测定,结果见表4,以此衡量该方法的精密度。

表4 方法精密度(n=8)

元素 分析波长/

nm

级次 GBW1307 U71Mn
标准值 测定值 RSD/% 标准值 测定值 RSD/%
Si 251.612 133 0.07 0.0685 0.412 0.243 0.2391 0.501
288.158 116 0.07 0.0692 0.365 0.243 0.2445 0.250
Mn 293.306 114 1.48 1.465 0.471 1.22 1.223 0.133
293.930 114 1.48 1.479 0.204 1.22 1.218 0.291
P 178.287 188 0.036 0.0355 0.520 0.014 0.0135 0.632
213.618 157 0.036 0.0362 0.447 0.014 0.0150 0.289
V 309.311 133 0.025 0.0248 0.486 0.012 0.0116 0.188
310.230 108 0.025 0.0249 0.219 0.012 0.0116 0.318
Ti 334.941 100 0.023 0.0231 0.291 0.0034 0.0032 0.345
336.121 100 0.023 0.0229 0.624 0.0034 0.0033 0.418
Al 309.271 108 0.029 0.0288 0.474 0.0158 0.374
396.152 85 0.029 0.0285 0.282 0.0165 0.206
Cu 324.754 103 0.035 0.0349 0.271 0.078 0.0776 0.311
327.396 102 0.035 0.0338 0.208 0.078 0.0787 0.246
Co 228.616 147 0.0051 0.598 0.038 0.0395 0.405
Ni 227.021 148 0.019 0.0192 0.691 0.029 0.0301 0.634
341.476 98 0.019 0.0179 0.279 0.029 0.0285 0.450
Cr 283.563 118 0.037 0.0369 0.220 0.016 0.0165 0.337
267.716 125 0.037 0.0365 0.430 0.016 0.0154 0.566
Mo 202.030 166 0.038 0.0378 0.414 0.0015 0.501
281.615 119 0.038 0.0382 0.370 0.0017 0.435
Mg 279.553 120 <0.001 0.706 <0.001 0.553
285.213 117 <0.001 0.901 <0.001 0.774

  从表4可见,相对标准偏差均小于1%,表明本方法的精密度良好。
4.4 方法的准确度
  为了评估该方法的准确度,对一系列标准样品进行测定,将测定结果与标准值进行对照比较,见表5。从表5可见,本方法的测定结果准确、可靠。

表5 标准样品测定结果对照表%

元素 分析波长/
nm
级次 GBW1301 GBW1304 GBW1309 GBW1311
标准值 测定值 标准值 测定值 标准值 测定值 标准值 测定值
Si 251.612 133 0.056 0.0556 0.290 0.294 1.11 1.12 0.685 0.695
288.158 116 0.056 0.0545 0.290 0.286 1.11 1.13 0.685 0.687
Mn 293.306 114 2.09 2.102 0.870 0.864 0.309 0.308 0.180 0.176
293.930 114 2.09 2.110 0.870 0.865 0.309 0.310 0.180 0.180
P 178.287 188 0.054 0.0535 0.011 0.010 0.012 0.011 0.018 0.018
213.618 157 0.054 0.0541 0.011 0.012 0.012 0.010 0.018 0.019
V 309.311 133   0.0022   0.0012 0.250 0.255 0.081 0.082
310.230 108   0.0021   0.0011 0.250 0.260 0.081 0.083
Ti 334.941 100   0.0004   0.0004 0.70 0.713 0.087 0.085
336.121 100   0.0003   0.0003 0.70 0.711 0.087 0.086
Al 309.271 108 0.015 0.0155 0.073 0.0721 0.186 0.187 0.036 0.035
396.152 85 0.015 0.0151 0.073 0.0761 0.186 0.186 0.036 0.039
Cu 324.754 103 0.200 0.198 0.034 0.0329 0.035 0.034 0.035 0.038
327.396 102 0.200 0.201 0.034 0.0332 0.035 0.033 0.035 0.035
Co 228.616 147   0.0051   0.0070   0.0065   0.0062
Ni 227.021 148 0.076 0.0756 0.290 0.305 0.021 0.021 0.020 0.021
341.476 98 0.076 0.0785 0.290 0.285 0.021 0.019 0.020 0.022
Cr 283.563 118 0.105 0.110 0.760 0.756 0.013 0.014 0.020 0.022
267.716 125 0.105 0.104 0.760 0.764 0.013 0.012 0.020 0.019
Mo 202.030 166   0.0054   0.0002 0.485 0.483 0.280 0.287
281.615 119   0.0061   0.0006 0.485 0.489 0.280 0.279
Mg 279.553 120   <0.001   <0.001   <0.001   <0.001
285.213 117   <0.001   <0.001   <0.001   <0.001
元素 分析波长/
nm
级次 GBW1312 GBW1322 GBW1327 GBW1504-1
标准值 测定值 标准值 测定值 标准值 测定值 标准值 测定值
Si 251.612 133 0.460 0.457 1.50 1.51 0.204 0.199 0.0146 0.0135
288.158 116 0.460 0.458 1.50 1.49 0.204 0.198 0.0146 0.0140
Mn 293.306 114 0.396 0.385 0.230 0.226 1.90 1.92 0.400 0.399
293.930 114 0.396 0.387 0.230 0.219 1.90 1.92 0.400 0.389
P 178.287 188 0.054 0.053 0.013 0.012 0.038 0.036 0.0064 0.0055
213.618 157 0.054 0.054 0.013 0.013 0.038 0.037 0.0064 0.0057
V 309.311 133 0.051 0.050 0.050 0.049 0.283 0.285   0.0003
310.230 108 0.051 0.051 0.050 0.048 0.283 0.286   0.0001
Ti 334.941 100 0.041 0.040 0.247 0.243 0.064 0.065   0.0066
336.121 100 0.041 0.040 0.247 0.241 0.064 0.065   0.0066
Al 309.271 108 0.028 0.026 0.066 0.069 0.276 0.278 0.014 0.011
396.152 85 0.028 0.027 0.066 0.066 0.276 0.273 0.014 0.012
Cu 324.754 103 0.030 0.029 0.320 0.316 0.116 0.113 0.085 0.084
327.396 102 0.030 0.028 0.320 0.315 0.116 0.114 0.085 0.085
Co 228.616 147   0.0048   0.0061   0.0039   0.0217
Ni 227.021 148 0.016 0.016 0.078 0.080 0.413 0.414 0.028 0.027
341.476 98 0.016 0.015 0.078 0.081 0.413 0.409 0.028 0.026
Cr 283.563 118 0.0095 0.0096 0.473 0.472 0.018 0.017 0.0053 0.0060
267.716 125 0.0095 0.0092 0.473 0.469 0.018 0.019 0.0053 0.0055
Mo 202.030 166 0.110 0.112   0.0048   0.0028   0.0059
281.615 119 0.110 0.109   0.0038   0.0025   0.0057
Mg 279.553 120   <0.001   <0.001   0.0005   <0.001
285.213 117   <0.001   <0.001   0.0004   <0.001

5 结论

  以同时测定碳钢、低合金钢中Si,Mn,P,V,Ti等十二个元素的分析方法为基础,对此类以CID作检测器的新型全谱直读等离子体光谱仪的分析性能进行探讨,找到了此类仪器测定钢铁样品的最佳分析测试参数并验证了仪器卓越的性能。特别是解决了以往仪器设备之不足,能够满足对P的分析测试,使其可取代化学法测P,提高了分析速度、降低分析成本。所建立的十二元素分析方法简便易行、快速准确、精密度好。

作者单位:攀枝花钢铁研究院

参考文献
  1 张展霞,刘志红.电荷转移器件在光谱学中的应用.现代仪器使用与维修,1995(1):4~6.
  2 中国光学学会光谱学会编译.电感耦合等离子体发射光谱图册.1986.

相关文章
文章录入:admin  收藏本页 关闭窗口 返回顶部
友情连接联系我们汇款信息热管设备热管教材热管论文常用软件网站地图
 

哈尔滨晨怡热管技术有限公司  哈尔滨晨怡网景电子有限公司热管分公司   地址:哈尔滨市南岗区南通大街256号
电话:0451-82589558 82589508 82589538 传真:0451-82552085  技术支持:13704813968  邮编:150001
电子信箱mail:rg@nx8.net  heat-pipe@hotmail.com
   heat.pipe@yahoo.com.cn
本站永久域名:  http://china-heatpipe.net  http://nx8.net   http://heatpipe.net.cn  http://rg.nx8.net
信息产业部ICP/IP地址信息备案:黑ICP备07500228    公安局国际联网备案登记:哈公网监备2301001165 Powered By: ACMS Version:2.0